Price Checker Manuals QuickQuotes About Terra Contact Us
NITROWATCH® System
 

NITROWATCH® System
NITROWATCH® System NITROWATCH® System NITROWATCH® System
Continuous low flow Intermittent high flow Gas bleed though RB valves
Internal humidity monitor
Purge control

NitroWatch®: Humidity Monitor & Control (Dual Purge System, RB valve required)
 
Specifications
Controller  
Dimensions: 11"W x 6.5"D x 3"H
Weight: 4 lbs., 7 ozs.
Power: 12 VDC (provided by Dual Purge™ System)
Case Material: 304 Stainless Steel
Output: 0-5 V
Measuring Range: 0-100%RH
Display Resolution:> ±.1%RH
Temperature Dependence: ±.04%RH/°C
Sensor  
Dimensions: 1.5" x .75" x 4"
Weight: 3 ozs.
Measuring Range: 0-100%RH
Accuracy (at 20°C): ± 2%RH
Housing Material: 304 Stainless steel
Calibration: None required
NitroWatch
Cat.# $/Each
Controller 9500-00A Login
Sensor
(One required per desiccator)
9500-02A Login

 
 
 
 
 
 
 
 

NITROWATCH® System
For operation with Terra's Dual Purge™ System

  • Automatically controls Dual Purge™ System to maintain the desired low relative humidity level (0%RH to ambient) in a desiccator or other enclosure
  • Reduces nitrogen costs by up to 78%
  • Continuously monitors and displays relative humidity level
  • Automatically compensates for shifting environmental variables, eliminating guesswork and reliance on supervisory personnel
  • New miniaturized sensor eliminates dead space inside glove boxes
  • Easy-program control functions

Purging a cabinet with nitrogen is an effective way to maintain a low-humidity storage environment. Using a flowmeter to control the nitrogen purge gas, however, leads to unnecessary expenses!

If you set the gas flow level too high, expensive nitrogen is wasted and cabinet seals are exposed to excessive pressure that could lead to leaks (or even ruptures!). If you set the flow level too low, you may wind up with ineffective contamination and moisture control and expensive component damage. Both problems have cost the semiconductor industry millions of dollars annually.

Terra Universal's NitroWatch® solves these problems by sensing and displaying the relative humidity of an enclosed environment, such as a desiccator, Faraccator, or glove box. More importantly, it operates with the Dual Purge™ System to precisely control the flow of nitrogen into the enclosure and maintain a preset humidity level.

Operation is simple and reliable. Simply program the desired level, and then let the system do its job. The fast-response, capacitive-type sensor detects the relative humidity; the level is indicated on a 4-digit LED display. The NitroWatch® controls Dual Purge System operation to maintain the setpoint level—no matter how often doors are opened, or how large the system is, or what the ambient humidity is, or what time of the day or night. High-flow purges are initiated only when a door is opened or the humidity climbs above the set-point. The moment acceptable conditions are restored, the system returns to low-flow purging.

Cuts Nitrogen Waste for Additional Savings!
Higher quality control and reduced part rejects are two convincing reasons to consider the NitroWatch®. But they aren't the only ones.

Your biggest savings will probably come in reduced nitrogen expenses. Typically, the nitrogen flow in a desiccator cabinet is set at a level high enough to compensate for access to stored parts during working hours; this means that a constant high-level purge is required. Yet in most cases, doors remained closed for over 75% of the time!

During off hours, weekends and holidays, unnecessarily high amounts of nitrogen are fed through the system and bled out as waste—even though no flow (or only a very minute flow) may be needed to maintain the appropriate internal humidity and positive pressure. Excess nitrogen bleeding can even endanger worker safety if ambient oxygen levels fall too low.

When you turn out the lights and lock the door, the NitroWatch® will cut out the nitrogen waste. This savings alone in most cases pays for the system within six months. Test data show typical savings of 78% of nitrogen consumption. And you'll be assured that stored components (and your workers) are safe.

Space-Saving Sensor
The NitroWatch® sensor is installed in the bottom desiccator chamber without interfering with shelves, doors, process equipment, or stored materials. This sensor features miniaturized components that provide better than 2%RH accuracy over the 10%-90%RH range. Further, it never requires calibration.

The NitroWatch® controller comes complete with all necessary controls for monitoring the relative humidity level. For automatic humidity control, the unit must be connected to Terra's Dual Purge System. Order the NitroWatch® Sensor separately (see below).

 

 
Preferred Client
Services Include:
QuickStart Manuals
Price Checker
Update Profile
You are not logged in.
Type email address and click Login.
Request a Quote
The QuickQuote planning form makes it easy to specify the system for your requirements
Product Selection


Desiccators
Section Resources
Plastic Desiccators
Stainless Steel Desiccators
Controllers & Accessories
Dual Purge N2 Controller
NitroWatch® Humidity Controller
Automatic RB™ Valve
Flowmeters
Gas Distributor
Gas Dryer
Nitrogen Generator
Disposable Gas Filters
DesiTray Renewable Drying System
Desiccant Boxes
Ground Continuity Indicator
Ionizing
Storage Racks & Latches
Gaskets
Casters


All prices and specifications are subject to change without notice. Crating charges apply to selected items.
  Terms and Conditions
  Precision Fabrication to Your Specs! Call to discuss your requirements for CNC routing and drilling.
  Top of page
 
 

You may need more than one relief/bleed valve to prevent overpressures in your desiccator – review
selection guidelines

 
     
Did You Know  
system integration
 

Terra is a comprehensive manufacturer who offers complete system integration, including electrical certification.